24 en | Programming steps Configuration Manager
970.129b | V2 | 2014.05 User Manual Bosch Security Systems
63. Acoustical feedback for technical failures
The Carephone 61/62 can be set to announce technical failures through
the LED lamps and the loudspeaker, or through the LED lamps only.
– 0 = loudspeaker off
– 1 = loudspeaker on. This is the default value.
– 2 = loudspeaker on from 7:00 to 21:00
– 3 = an acoustical failure indication (message or beep) is repeated with
a pause of 10 seconds until the Action button is pressed.
– 4 = an acoustical failure indication (message or beep) is repeated
during day time (from 7:00 to 21:00) with a pause of 10 seconds until
the Action button is pressed.
Settings 2 and 4 are only available if the time and date are set. This can be
done through keyboard programming, Configuration Manager with serial
serial connection or via a remote date/time update from a receiver.
77. Pre-alarm time
The pre-alarm time of the Carephone 61/62 can be programmed. This is
the time frame in which an emergency call can still be stopped.
– Select a setting, in steps of 10 seconds, between 0 and 6.
– 0 = off, 1 = 10 s, 2 = 20 s, etc. Default value is 1.
24. Confirmation with call
When the repeated emergency call has been locally acknowledged, you can
speak directly to the monitoring centre.
27. Presence marking - Service done
The presence marking function allows staff members to mark their
presence or signal that the service is done by pressing the daily button.
This is handled without a speak/listen connection and does not require an
answer by the operator. When the presence marking is activated, the
destination numbers to be called are set in the call sequence of step 58 of
the tab Function assignment. This function resets the activity monitor,
which must be set. See step 40 of the tab General.
02. CPC Adaptation
You can set the Carephone 61/62 to use a specific CPC adaptation when
the CPC protocol is used. Check the tick-box to enable this function.